Search found 455 matches

by jry
Wed Jan 07, 2015 10:55 am
Forum: Test Equipment & RF Equipment Alignment
Topic: Aeroflex (IFR) Questions
Replies: 629
Views: 295313

Re: Aeroflex (IFR) Questions

the option field on the configuration report shows the software option field. Usually this is 00000002 which has the file system enabled. Believe the rest of the bits there are for cellular protocols or options...not really that useful anymore IMHO. Bottom line is that the option bits do not have to...
by jry
Tue Jan 06, 2015 2:38 pm
Forum: Test Equipment & RF Equipment Alignment
Topic: Aeroflex (IFR) Questions
Replies: 629
Views: 295313

Re: Aeroflex (IFR) Questions

but chances are it's something external in your setup if the main display is working OK
by jry
Tue Jan 06, 2015 11:28 am
Forum: Test Equipment & RF Equipment Alignment
Topic: Aeroflex (IFR) Questions
Replies: 629
Views: 295313

Re: Aeroflex (IFR) Questions

always active. Plate is there to protect the contacts from dirt and damage.
There is a cable internal that may have gotten disconnected or you have an issue on the video card
by jry
Mon Dec 22, 2014 2:45 pm
Forum: Test Equipment & RF Equipment Alignment
Topic: HP8920A locking up
Replies: 2
Views: 2065

Re: HP8920A locking up

re-cap the power supply ...at least the low voltage filter section.

The 8920A is in the twenty year old range and the power supplies are starting to go ....often preceded by a bad smell or puffs of smoke on power up
by jry
Fri Nov 21, 2014 7:09 am
Forum: Test Equipment & RF Equipment Alignment
Topic: R2600 going straight to CAL mode.
Replies: 5
Views: 2525

Re: R2600 going straight to CAL mode.

COMP-18-3-NMH is what I use
by jry
Thu Nov 20, 2014 2:15 pm
Forum: Test Equipment & RF Equipment Alignment
Topic: R2600 going straight to CAL mode.
Replies: 5
Views: 2525

Re: R2600 going straight to CAL mode.

processor battery dead
by jry
Thu Nov 13, 2014 6:18 am
Forum: Test Equipment & RF Equipment Alignment
Topic: Dulplexer question
Replies: 66
Views: 16208

Re: Dulplexer question

That rusty pipe you are bolted to may not help much with the noise level. Google rf interference and rust

You may want to find something else to mount that antenna to.
by jry
Tue Oct 21, 2014 2:33 pm
Forum: Test Equipment & RF Equipment Alignment
Topic: Difference in R2001A and R2001C
Replies: 14
Views: 3858

Re: Difference in R2001A and R2001C

The C version operations ( not sure its any different that the A ) was in the full service manual
by jry
Fri Oct 10, 2014 2:21 pm
Forum: Test Equipment & RF Equipment Alignment
Topic: Aeroflex (IFR) Questions
Replies: 629
Views: 295313

Re: Aeroflex (IFR) Questions

on the digital card repair just the aluminum electrolytics should need to be replaced. The issue is that they will eventually fail if they are not leaking and defective now. On the Watch Dog timer chip replacement on the processor it takes a $600 XLTEK programmer ( at least that what I spent ) that ...
by jry
Wed Oct 08, 2014 12:05 pm
Forum: Test Equipment & RF Equipment Alignment
Topic: COM120B Real time clock failure
Replies: 1
Views: 1704

Re: COM120B Real time clock failure

you double posted this ...replied in the other thread
by jry
Wed Oct 08, 2014 12:04 pm
Forum: Test Equipment & RF Equipment Alignment
Topic: Aeroflex (IFR) Questions
Replies: 629
Views: 295313

Re: Aeroflex (IFR) Questions

There is a Dallas RTC chip on the processor board next to the DRAM battery ( more or less ) Is has an internal battery but not sure that would cause the issue you are seeing but it's possible. I have replaced these before and they have to be programmed as well ...usually install a socket and a dab o...
by jry
Tue Oct 07, 2014 1:40 pm
Forum: Test Equipment & RF Equipment Alignment
Topic: Aeroflex (IFR) Questions
Replies: 629
Views: 295313

Re: Aeroflex (IFR) Questions

The digital card is in the first slot so you can remove the front of the 120 to get full access. You can remove the EL from the front and lay on top but given you have no useful information on the screen would just leave it disconnected for the moment. I would make sure you have 1. removed all the a...
by jry
Fri Oct 03, 2014 9:59 am
Forum: Test Equipment & RF Equipment Alignment
Topic: R2600B/HS with CRT problem
Replies: 4
Views: 2436

Re: R2600B/HS with CRT problem

carefully with a very small screwdriver and insulated handle and a jumper clip ..utube video on this
by jry
Fri Oct 03, 2014 6:22 am
Forum: Test Equipment & RF Equipment Alignment
Topic: R2600B/HS with CRT problem
Replies: 4
Views: 2436

Re: R2600B/HS with CRT problem

Boy this is going to cost a fortune to repair 8) I think all you need to do is CLEAN the CRTespecially by the anode and the HV cable to the flyback, I often tape the Metal chassis (wide clear packing tape ) by the anode connection just a precaution. I would dis-assemble the front from the unit unplu...
by jry
Wed Sep 17, 2014 8:41 am
Forum: Test Equipment & RF Equipment Alignment
Topic: Another IFR1600S Question
Replies: 10
Views: 3135

Re: Another IFR1600S Question

the lead set for the 1600 can use a standard uncovered banana plug. Typically use a standard Banana to BNC adapter on those inputs Believe the 1600/1900 lead set was initially from Coline and had a retractable banana plug cover but was deeper insertion than the standard Coline set that was used with...
by jry
Sat Sep 13, 2014 10:16 am
Forum: General Motorola Solutions & Legacy Radio Discussion
Topic: XTS3000 - CPS Read Block Help
Replies: 7
Views: 2820

Re: XTS3000 - CPS Read Block Help

other option would be to verify your setup with another radio. If you are getting a consistent error ( fails at the same point every time )on the codeplug unpack from that radio it could very well be that the codeplug has become corrupt. Usually the controller board is what gets replaced on half the...
by jry
Thu Sep 04, 2014 6:50 am
Forum: General Motorola Solutions & Legacy Radio Discussion
Topic: XTS3000 - CPS Read Block Help
Replies: 7
Views: 2820

Re: XTS3000 - CPS Read Block Help

that is the correct and latest SW for an XTS3000. Is there an encryption module installed ? Just wondering if the codeplug has encryption configured but the module was renoved. May be irrelevant. If you go into test mode ( hit the bottom side button five times ) and see if it reports the encryption ...
by jry
Sun Aug 24, 2014 11:22 am
Forum: Test Equipment & RF Equipment Alignment
Topic: IFR1200 Super S
Replies: 4
Views: 3102

Re: IFR1200 Super S

You really already have all the hardware for the tracking gen ...that does come standard on the Super S with the exception of the amp for the high level output and a short SMB jumper between the Duplex Track module and the first spec anal module The feature is enabled on the processor card ( same on...
by jry
Sat Aug 23, 2014 9:14 am
Forum: Test Equipment & RF Equipment Alignment
Topic: IFR1200 Super S
Replies: 4
Views: 3102

Re: IFR1200 Super S

The tracking gen is an option even on the 1200 Super S .

It's a feature normally enabled at the factory in the processor card.

There is normally an additional high level amp installed for the tracking gen output with the Option 12 as well
by jry
Mon Aug 11, 2014 5:07 am
Forum: Test Equipment & RF Equipment Alignment
Topic: HP8591A Counterlock (A25) Troubleshooting
Replies: 2
Views: 1818

Re: HP8591A Counterlock (A25) Troubleshooting

Steve,

You may want to try the HP AGiLENT group on Yahoo ...someone there may have some input.

My 8591A has always worked flawlessly
by jry
Sun Aug 10, 2014 2:36 pm
Forum: Test Equipment & RF Equipment Alignment
Topic: COM-120A/B ...Bug or broken ?
Replies: 0
Views: 1941

COM-120A/B ...Bug or broken ?

Was working on a COM-120A that has had the controller/processor upgraded to the B version with main firmware at 4.12 IIRC When doing the DVM calibration the screen and procedure in the manual shows looking for 1 ,10 and 100VAC along with 2VDC. During the cal it will fail everything but the 100VAC us...
by jry
Wed Jul 30, 2014 8:04 am
Forum: Test Equipment & RF Equipment Alignment
Topic: IFR-1200S Off Frequency
Replies: 16
Views: 12878

Re: IFR-1200S Off Frequency

You may want to replace Q15 and C69 ( common 2n3905 and 100pf ) and see if you can get the freq above 75.2 with the MV209 varactor out of the circuit first. The worst case is that the crystal has aged off frequency the effective load capacitance will need to be decreased or the crystal replaced. ( I...
by jry
Sat Jul 26, 2014 4:33 am
Forum: Test Equipment & RF Equipment Alignment
Topic: IFR-1200S Off Frequency
Replies: 16
Views: 12878

Re: IFR-1200S Off Frequency

i believe you also posted the spectrum analyzer is off as well so it seems pretty certain from everything the low loop output is off frequency. I would be good to know the exact amount off since that may help determine which divider may be suspect. I would pull the low loop module and remove it from...
by jry
Fri Jul 25, 2014 6:54 am
Forum: Test Equipment & RF Equipment Alignment
Topic: IFR-1200S Off Frequency
Replies: 16
Views: 12878

Re: IFR-1200S Off Frequency

Sounds like you have an issue in the low loop. Could also be an issue loading the correct values into the low loop from the uP From the service manual info you can easily get the correct low loop frequency for the current RF setting and measure the output form the LOW LOOP module. The low loop is wh...
by jry
Thu Jul 17, 2014 4:39 am
Forum: Test Equipment & RF Equipment Alignment
Topic: extender card for R2000 series .
Replies: 4
Views: 3026

Re: extender card for R2000 series .

I do have an extender kit for the R2000 series ....missing one of the extenders though.
Which size extender do you need ? there were two in the kit.
by jry
Mon Jul 07, 2014 7:16 am
Forum: Test Equipment & RF Equipment Alignment
Topic: motorola 2600b
Replies: 2
Views: 1798

Re: motorola 2600b

I would start by pulling the boards and looking for leaky SMT electrolytic capacitors. There is a million screws and the card guides can be pretty stiff as well. It basically amounts to a complete tear down of the unit and inspection. It will take time and look closely. If you have a capacitance met...
by jry
Mon Jun 30, 2014 6:34 am
Forum: Test Equipment & RF Equipment Alignment
Topic: R2200 Service Manual or schematic
Replies: 5
Views: 2400

Re: R2200 Service Manual or schematic

Check the Repeater Builders site the manual is there under the motorola test equipment. The fuse should be a standard 1/16 amp ...same as the R2600. Search the board since I think the part generic number has been posted. Also check the manual t make sure that value is correct. You cold also use a 4 ...
by jry
Tue Jun 17, 2014 4:57 pm
Forum: Test Equipment & RF Equipment Alignment
Topic: Transmitter output values gm1200
Replies: 4
Views: 2137

Re: Transmitter output values gm1200

The CPS/RSS values vary from radio to radio as to what impact they have. Motorola was never consistent on that. Normally I would just align per the manual and adjust down from there for the actual operating power levels you want. The radio is designed for a wide frequency operation no matter what yo...
by jry
Tue Jun 17, 2014 11:25 am
Forum: Test Equipment & RF Equipment Alignment
Topic: Transmitter output values gm1200
Replies: 4
Views: 2137

Re: Transmitter output values gm1200

There are two different things you are talking about but they are related. One is the power calibration level at the various frequency which essentially provides a flat output across the rated frequency range. The second is the power out adjustment ...usually set first and set at the mid-point frequ...
by jry
Tue Jun 03, 2014 10:07 am
Forum: Test Equipment & RF Equipment Alignment
Topic: R-2008 Load Resistor
Replies: 2
Views: 1722

Re: R-2008 Load Resistor

The other choice in looking at the 2001D input assembly would be to replace both the input attenuators. Total needs to be 30db for the RF gen and signal level to be correct on the I/O port. USe a 10db High power and a 20 db lower power and re-cal the wattmeter since the power tap will be 4 db higher...
by jry
Tue Jun 03, 2014 5:06 am
Forum: Test Equipment & RF Equipment Alignment
Topic: R-2008 Load Resistor
Replies: 2
Views: 1722

Re: R-2008 Load Resistor

Not sure if you are referring to a C or a D version R2008 If it's a D version than I believe that's a 14db 120 Watt attenuator. Last time I replaced one had to use a combination of a 10db 100Watt and a 4db 10 Watt . Required a mod of the watt meter board to get the two attenuators to fit and mount. ...
by jry
Tue May 27, 2014 2:35 pm
Forum: Test Equipment & RF Equipment Alignment
Topic: IFR-1600 WITH PHASE LOCK PROBLEMS
Replies: 30
Views: 7727

Re: IFR-1600 WITH PHASE LOCK PROBLEMS

May be good to understand the current issues since it sounds like you may have fixed one of the problems. Which main self-test is failing first and what are the current extended test results ? If you have not done so already verify the power supply levels and re-seat the PCB's and verify the RF conn...
by jry
Mon May 12, 2014 9:27 am
Forum: Test Equipment & RF Equipment Alignment
Topic: IFR-1600 WITH PHASE LOCK PROBLEMS
Replies: 30
Views: 7727

Re: IFR-1600 WITH PHASE LOCK PROBLEMS

you may have multiple issues or a bad RF I/O board ? the 1st LO gen passes but the 1st lo rec fails. ...the only difference is the buffered ref signal going in and the data programming between those two ...identical circuits otherwise. One would assume the buffered reference is good but you could sw...
by jry
Fri May 09, 2014 5:29 am
Forum: Test Equipment & RF Equipment Alignment
Topic: IFR-1600 WITH PHASE LOCK PROBLEMS
Replies: 30
Views: 7727

Re: IFR-1600 WITH PHASE LOCK PROBLEMS

1. Remove A22 RF I/O (para. 2-2-4) and install on PC Board Assembly Extender in Test Set. 2. Run Self Test 15 in Extended Mode until failure is detected. 3. Verify voltage for failed test with DMM at A22P2 (Figure C-22). Failed Test Location Voltage If incorrect, perform 15.1 A22P2, Pin 11 B 5 Vdc (...
by jry
Tue May 06, 2014 2:17 pm
Forum: Test Equipment & RF Equipment Alignment
Topic: IFR-1600 WITH PHASE LOCK PROBLEMS
Replies: 30
Views: 7727

Re: IFR-1600 WITH PHASE LOCK PROBLEMS

My guess is you have the 2nd LO 1Mhz out ( 10Mhz ref divide by 10 ) missing. Maybe a bad divider or buffer amp in that module Check for the 1Mhz signal at the 3rd LO and or 90Mhz gen modules. The 1Mhz should be one of the SMB cables going into the top of the modules ( come from the divider output th...
by jry
Sun Apr 13, 2014 8:19 am
Forum: Test Equipment & RF Equipment Alignment
Topic: Motorola General Dynamics R2590 Manuals ?
Replies: 7
Views: 2785

Re: Motorola General Dynamics R2590 Manuals ?

Tried Motorola Online to see if they still had a copy of the manual but no joy.

Is the resource CD you are referring to the one for the 2600 series or one that covered both the 2590 and 2600 ?

I have the 2600 resource CD but have not tried any of the utilities to see if they work with the 2590.
by jry
Sun Apr 06, 2014 11:26 am
Forum: Test Equipment & RF Equipment Alignment
Topic: Motorola General Dynamics R2590 Manuals ?
Replies: 7
Views: 2785

Re: Motorola General Dynamics R2590 Manuals ?

believe the main application version was listed as 16.1 ...doubt the boot loader and graphics versions are that important
by jry
Sun Apr 06, 2014 4:51 am
Forum: Test Equipment & RF Equipment Alignment
Topic: Squelch on MOTOROLA R 2600
Replies: 3
Views: 2045

Re: Squelch on MOTOROLA R 2600

could you be a little more descriptive on what that means ?
Not sure if it means the squelch is stuck open or you get no RX audio.
There is an LED on the front panel on top of the squelch control to indicate if the squelch is open or closed internally at least
by jry
Sun Apr 06, 2014 4:46 am
Forum: Test Equipment & RF Equipment Alignment
Topic: Motorola General Dynamics R2590 Manuals ?
Replies: 7
Views: 2785

Re: Motorola General Dynamics R2590 Manuals ?

Thanks for the info.
The manuals I found still had the red preliminary across them.
I will check to see what version is in this unit on the firmware.
by jry
Wed Apr 02, 2014 5:55 am
Forum: Test Equipment & RF Equipment Alignment
Topic: Motorola General Dynamics R2590 Manuals ?
Replies: 7
Views: 2785

Motorola General Dynamics R2590 Manuals ?

Anyone have a scan or pdf of the R2590 Manuals ?

Tried General Dynamics but no response and all the ones I found on the web are just the preliminary Ops manual.

Last firmware would be great as well if anyone kept a copy.

Thanks !
by jry
Mon Mar 10, 2014 5:48 am
Forum: Test Equipment & RF Equipment Alignment
Topic: R2600B vs R2008D
Replies: 5
Views: 2305

Re: R2600B vs R2008D

response is not always flat depending on the frequency being used as well...always used a jumper first to see what flat response looked like, Usually provided a decent visualization of the response curve but actual insertion loss and rejection values would have to be determined at the point frequenc...
by jry
Thu Mar 06, 2014 8:59 am
Forum: Test Equipment & RF Equipment Alignment
Topic: R2600B vs R2008D
Replies: 5
Views: 2305

Re: R2600B vs R2008D

not really supported ...just turn on the duplex gen with 0 offset ...only tried it on D series
by jry
Thu Mar 06, 2014 6:08 am
Forum: Test Equipment & RF Equipment Alignment
Topic: R2600B vs R2008D
Replies: 5
Views: 2305

Re: R2600B vs R2008D

R2600B R200XD is easier to use due to the information on the front panel that you have to bounce around screens on the R2600 for though. The R2600 is smaller and more capable with an external display capability as well as having a better tracking gen. The self test is more sophisticated giving a hig...
by jry
Thu Mar 06, 2014 6:02 am
Forum: Test Equipment & RF Equipment Alignment
Topic: Motorola R2600 dim display
Replies: 8
Views: 2970

Re: Motorola R2600 dim display

If you have an old monitor that did CGA/EGA it would be certainly worth a shot . I use an XVGA ( Ebay search should find them ) adapter which is intended for industrial CRT replacement converters but other may work ...I would avoid ones intended for old gaming applications. You may have to look clos...
by jry
Wed Mar 05, 2014 1:52 pm
Forum: Test Equipment & RF Equipment Alignment
Topic: Motorola R2600 dim display
Replies: 8
Views: 2970

Re: Motorola R2600 dim display

that converter will not work with this unit. Have not looked recently but there may be an internal adjustment on the CRT assembly for the intensity. The CRT is a Motorola industrial TTL OEM 7" display IIRC...should be labeled but never did find and information on the series used. Believe there ...
by jry
Fri Feb 28, 2014 5:47 am
Forum: Test Equipment & RF Equipment Alignment
Topic: IFR1500 CRT Filament Voltage
Replies: 6
Views: 2466

Re: IFR1500 CRT Filament Voltage

The poor RX is typically someone keying into the RX port ...probably blew some of the components / attenuators. Not too bad to access since it's on the top shelf . Usually re-build the fixed attenuators with surface mount resistors of the correct value forthe required 20db --the cans can be opened a...
by jry
Fri Feb 07, 2014 6:02 am
Forum: Test Equipment & RF Equipment Alignment
Topic: IFR1500 CRT Filament Voltage
Replies: 6
Views: 2466

Re: IFR1500 CRT Filament Voltage

Other option would be to add the chopper daughter board and use the off the shelf transformer. The chopper is only a handful of components and a 1/8 aluminum L bracket for the transistor mounting and heatsink. Using SMT components may be a thought if going this route. Sounds like you have it either ...
by jry
Wed Feb 05, 2014 6:29 am
Forum: Test Equipment & RF Equipment Alignment
Topic: IFR1500 CRT Filament Voltage
Replies: 6
Views: 2466

Re: IFR1500 CRT Filament Voltage

The output voltage may depend slightly on the version CRT controller board. The transformer wasn't changed but the early version used a feedback winding to setup the primary oscillations. The later version used a fixed frequency chopper daughter board and no longer used the feedback winding. These l...
by jry
Mon Jan 06, 2014 3:54 pm
Forum: Test Equipment & RF Equipment Alignment
Topic: IFR 1200S Alignment & Firmware
Replies: 6
Views: 3055

Re: IFR 1200S Alignment & Firmware

As mentioned before would go through the whole 1200 calibration in order. That way you have a solid base ...would use the high loop adjustment procedure on the KG electronics website ...much more direct method of doing the alignments. The spectrum analyzer adjustments have quite a bit of interaction...