I would like to ask you about a problem occurred on my COM 120b.
In the following spans my analyser shows the track a little bit shifted away:
100KHz division trace is shifted
500Khz division trace is shifted
1MHz division trace is shifted
-----------------------------------------

-----------------------------------------
1KHz division is ok
2Khz division is ok
5KHz division is ok
10KHz division is ok
20KHz division is ok
50KHz division is ok
200KHz division is ok
-----------------------------------------


This problem could came out when I exchanged the old RTC DS1286 with
drained battery with a new one.
I carried out the whole calibration of the analyser on both hardware
and software parts as explained in the service manual, but the problem
is still present.
Should I ask you if your COM 120b have the same issue, considering
it's not so serious for my job?
Wowbagger what do you think about?
Maybe could be the filter used to plot those spans deriving a little
bit from the standards?
Do you know if exist some kind of software offset to be set via RS232
commands to fix this thing?
Thank you
Best regards,
Gianfranco